Octreoscan protocol
- Preparation
- Patient must follow preparation as outlined on octreotide scan guideline handout
- If patient is on short acting octreotide, they must be off of it for 24 hours
- If patient is on long acting octreotide, they must be off of it for 4-6 weeks.
- Isotope administration
- Patient will receive a 6mCi intravenous injection of In111 To assay the dose, use a dose calibrator using a copper dipper.
- Camera Parameters Planar imaging
- MELP collimators used for imaging
- 8 cm/min table speed
- Matrix: 256X1024
- Zoom: 1
- Camera peaked at In111 with a 15% window
- Camera Parameters SPECT
- MELP collimators used for imaging
- Camera peaked at In 111 with a 15% window with scatter window
- Matrix: 64X64
- Zoom: 1
- Time: 35 sec per stop
- Views: 32 per bed
- Detector configuration 180
- Orbit non circular
- Mode step/shoot
- Scanning protocol
- Patient will be scanned supine position and scanned, 4 and 24 hours post injection, from the superior aspect of the skull to the feet.
- 2 to 3 bed SPECT imaging with CT imaging for attenuation correction.
- Additional imaging as requested by radiologist.
- SPECT/CT Reconstruction
- Nuclear Medicine Parameters
- Iterative OSEM 2D (4 subsets, 8 iterations)
- Gaussian 6.0 filter
- Nuclear Medicine Parameters
- Chang’s Attenuation Coefficient 0.15
- CT Parameters
- Abd Window
- Body 5.0 B30s Med Smooth
- 5 mm Slice
- Post processing/Display/Achieve
- All acquired images will be adjusted for optimal viewing and appropriate labeling.
- All SPECT images will be displayed in coronal and sagittal planes with and without CT fusion and NM MIP
- Images will be sent to the PAC’s system and Syngo Via for archiving/physician review