Routine heart

ROUTINE HEART (rev. 7/1/2013)

1. 3 plane localizer

Heart at isocenter

2. Axial Bright Blood FISP

Usually 25-30 slices, cover from aortic arch through bottom of heart

Straight axial

Prescribe off localizer

Do not clip the bottom of heart 

Set to multiple breath holds (3) - tailor to patient's ability

Prospectively gated (Physio - capture cycle)

Can be OVERLAPPED for better images - set distance factor to -50%

3. Axial Dark Blood HASTE

Usually ~20 slices (thicker than FISP)

Straight axial

Same coverage as FISP

Free breathing

Prospectively gated (Physio - capture cycle)

4. CINE 2 Chamber

1 slice CINE, retrospectively gated (no need to capture cycle)

Breath hold

Prescribe off localizer or bright blood FISP

Center of plane in center of mitral valve

Rotate so it passes through apex of LV

Make perpendicular (Protocol - click "Perpendicular")

    *make sure the plane you want to be perpendicular to is "highlighted" in planning window

5. CINE FAKE SHORT AXIS

1 slice CINE, retrospectively gated (no need to capture cycle)

For planning purposes

Breath hold

Prescribe off CINE 2 Chamber - single short axis slice through mid LV 

Make perpendicular (Protocol - click "Perpendicular")

6. CINE 4 CHAMBER

1 slice CINE, retrospectively gated (no need to capture cycle)

Breath hold

Prescribe off BOTH CINE 2 Chamber and FAKE SHORT AXIS


    • FAKE SHORT AXIS - plane through most angled portion of RV and mid LV papillary muscle
    • 2 CHAMBER - plane through apex and mid mitral valve

7. T2 DARK BLOOD

For edema imaging if applicable - acute MI and infiltrative disease

Breath hold each slice

Prospectively gated (Physio - capture cycle)

  Short axis stack - 4 or  5 slices through LV spaced accordingly. Prescribe off CINE 4 CHAMBER and CINE 2 CHAMBER

  4 chamber - single slice, copy slice position from CINE 4 CHAMBER

  2 chamber - single slice, copy slice position from CINE 2 CHAMBER

8. FIRST PASS PERFUSION

3 short axis slices through LV spaced accordingly (one base, one mid LV, one apical) - prescribe off CINE 4 CHAMBER and CINE 2 CHAMBER

1 long axis slice through LV - plan as a 4 chamber view using FAKE SHORT AXIS and 2 CHAMBER

Inject 1/4 of double dose contrast

Start scan and inject contrast simultaneously (click both buttons at same time)

When contrast reaches RV, ask patient to hold their breath

They can slowly let air out when they can no long hold it

Make sure images are good with contrast and no wrap. If bad, adjust and repeat with another 1/4 dose of contrast. 

If good, inject remainder of contrast dose.

*Note TIME of injection - do TI scout 10 minutes post injection

9. CINE SHORT AXIS STACK

Most important sequence for quantification - cannot have motion!

Stack of CINE images through LV - breath hold, retrospectively gated (no need to capture cycle)

Prescribe off CINE 4 CHAMBER and CINE 2 CHAMBER - perpendicular cross sectional slices on both views

Cover from mitral valve through apex - can add or remove slices to ensure adequate coverage

**Check plane on axial and sagittal scouts to ensure no wrap

10. CINE 3 CHAMBER

3 chambers = LV, LA, AORTA

1 slice CINE, retrospectively gated (no need to capture cycle)

Breath hold

Prescribe off "snowman view" - one of the basal most CINE SHORT AXIS STACK slices - bisect the snowman - through LV and AORTA

Make perpendicular (Protocol - click "Perpendicular")

* Check position on 4 CHAMBER CINE - Through LV apex and LA

* alternative is to 3 point plan LV, LA, and aorta

11. TI SCOUT

1 short axis slice through LV - COPY slice position from CINE FAKE SHORT AXIS

Prospectively gated (Physio - capture cycle)

Free breathing

~10 minutes post contrast injection (does not have to be exact!)

12. MDE SINGLE SHOT

Copy slice positions from appropriate CINE sequences

Prospectively gated (Physio - capture cycle)

Find inversion time where myocardium is BLACK and blood is BRIGHT

* this should look like a delayed enhancement image

* tram tracking is TOO EARLY

Enter this inversion time + 20-30 msec (Contrast - TI)

Breath hold better, but free breathing okay

Short axis stack same number of slices as CINE

2, 3, 4 chamber - 5 slices each

13. MDE SEGMENTED

Copy slice positions from appropriate CINE sequences

Prospectively gated (Physio - capture cycle)

Enter inversion time from scout + 20-30 msec (Contrast - TI)

   *rule of thumb is to increase the TI ~30 msec for every ~5 mins of scanning post TI scout

Breath hold always - important to give good instructions as very motion sensitive

Short axis stack same number of slices as CINE

2, 3, 4 chamber only one slice

14. AORTA PHASE CONTRAST

Use AXIAL BRIGHT BLOOD FISP, find the slice with the right main pulmonary artery, right click and then click "COPY IMAGE POSITION"

Breath hold

Ensure no aliasing, otherwise increase VENC and repeat