Meckel’s Scan
Preparation: none
Dose:10mci TC99m
Procedure: An intravenous line will placed in the patients arm prior to being positioned under the camera. The patients will be positioned on the imaging table with the tip of the xiphoid at the top of the imaging field and the urinary bladder at the bottom of the imaging field. The patient should be centered in the middle of the imaging table. Detector 1 will placed above the imaging table close to the patient’s abdomen.
Number of frames: 63
Time per frame: 60sec
Matrix 128×128 (both segments)
Zoom:1
Detector:1
Analyzer: TC99m-NMG
Orientation: head out
Images will be adjusted for optimal viewing intensity and sent to PACs.