Liver Spleen Study
- Preparation
- No patient preparation
- Isotope administration
- Patient will receive 5 mci Tc99m sulfur colloid intravenously
- Camera Parameters
- Planar imaging
- LEHR collimators
- Matrix 128X128
- Zoom: 1.23
- Camera Preset: Tc99m-NMG
- Detector: both
- 500 Kcts per view
- SPECT imaging
- LEHR collimators used for imaging
- Camera peaked at Tc99m with a 15% window with scatter window
- Matrix: 128X128
- Zoom: 1
- Time: 25 sec per stop
- Views: 64
- Rotation: clockwise
- Start angle: 0
- Degree rotation: 180
- Detector configuration 180
- Orbit non circular
- Mode step/shoot
- Patient position: supine
- Orientation: head out
- Camera preset: Tc99m-MNG
- Scanning protocol
- Planar Imaging
- Patient will be injected with radiotracer intravenously
- Patient will wait 20 minutes before scanning to begin
- Immediately prior to placing patient on scanning table the patient should be asked to void.
- All metal objects should be removed from patient prior to scan start
- Patient will lay supine on the table with the head facing out.
- Patient will be scanned with liver and spleen centered under the camera
- The static images to be taken are Anterior, Anterior with liver markers, posterior, right lateral, left lateral, LAO, RPO
- Additional static views will be acquired for areas of interest or at the request of the radiologist supervising the exam.
- SPECT Imaging
- Patient will be injected with radiotracer intravenously
- Patient will wait 20 minutes before scanning to begin
- Immediately prior to placing patient on scanning table the patient should be asked to void.
- All metal objects should be removed from patient prior to scan start
- Patient will lay supine on the table with the head facing out.
- Patient will be scanned with liver and spleen centered under the camera
- 1 bed SPECT imaging with CT imaging for attenuation correction.
- Additional imaging as requested by radiologist.
- SPECT/CT Reconstruction
- Nuclear Medicine Parameters
- Filtered backprojection
- Butterworth cutoff of .45 and order of 5
- Chang’s Attenuation Coefficient 0.15
- Automatic contours
- Projection angel is 2
- Edge strength of .5
- Background threshold of 5
- Kernel 3X3
- CT Parameters
- Bone Window
- Body 5.0 B30s Med Smooth
- 5 mm Slice
- Post processing/Display/Achieve
- All acquired images will be adjusted for optimal viewing and appropriate labeling.
- Planar images will be sent to the PAC’s system
- All SPECT images will be displayed in coronal and sagittal planes with and without CT fusion and NM MIP, and sent to the PAC’s system and Syngo Via for archiving/physician review