SPHP Radiology Protocols

Radiology Protocols

Noncompaction

Noncompaction (rev. 06/23/2023)

Indication: Trabeculated ventricle on echo, evaluate for LV noncompaction

SEQUENCES

Localizer

Axial bright blood MBH chest

2 chamber CINE stack x 3

Fake short axis CINE

4 chamber CINE stack x 3

T2 dark blood SA

T2 dark blood 2 chamber

T2 dark blood 4 chamber

— CONTRAST–

optional 4 slice first pass perfusion

Short axis stack CINE

3 chamber CINE stack x 3

Aorta through plane Phase Contrast (VENC 150)

TI scout

MDE single shot SAS

MDE single shot 2 chamber

MDE single shot 4 chamber

MDE single shot 3 chamber

** If filling defect in LV (possible thrombus):

 

MDE single shot SAS long TI

MDE single shot 2 chamber long TI

MDE single shot 4 chamber long TI

MDE single shot 3 chamber long TI

**optional segmented and 3D MDE

NOTES

This is routine heart with T2, with minor changes:

 

2 chamber, 4 chamber, and 3 chamber CINEs are done as a stack of THREE slices, with 50% spacing.

 

If visible filling ventricular mass on T2 or CINE, then repeat MDE single shot sequences with long TI of 800-1000 to evaluate for THROMBUS